Model Number UHI-4
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2023
Event Type  Injury  
Manufacturer Narrative
To date, this device has not been returned for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
 
Event Description
The customer reported to olympus that during preparation for laparoscopic procedure, this high flow insufflation unit turned on for 0.5 seconds and then turned off with a beeping sound, making it unusable.The patient was taking warfarin and had chosen a laparoscopic procedure, which could cause as little bleeding as possible.However, since the device could not be used, the procedure was changed to an open surgery.The device, as reported, had not been used for over a year.There were no reports of further patient or user harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the customer¿s reported complaint was confirmed.In addition, per the log analysis, e03 was confirmed, and the pressure sensor was broken.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the reported event occurred due to a faulty pressure sensor on the main board.However, the root cause could not be determined.This supplemental report includes a correction to b5 and g2 to provide information that was inadvertently not included on the initial medwatch.An update has been made to d9 and h3 from the initial medwatch.Also, new information has been added to d8.Olympus will continue to monitor field performance for this device.
 
Event Description
The following additional information was received: it was confirmed that the laparotomy procedure was completed successfully, and there was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16729125
MDR Text Key313188836
Report Number3002808148-2023-03662
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-