Model Number 429878 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/30/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during the implant procedure of the left ventricular (lv) lead, two guidewires were attempted to be used to place the lead and a third guidewire was utilized.The third guidewire would not advance beyond the end of the lv lead and became stuck inside the lead.The physician was not able to advance or remove the guidewire.The lv lead was attempted and not used.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the full lead was returned and analyzed.Analysis indicated the distal conductor of the lead was obstructed due to a competitor guidewire stuck in the lumen.The competitor guidewire was kinked/buckled.Visual analysis of the lead indicated damage at implant.The analyst noted no anomalies were found on the lead.The competitor guidewire's distal tip was kinked which caused the distal end of the lead to be obstructed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|