ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ NAVLINK MODULE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H420995
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2023
Event Type  malfunction  
Event Description
During a pulmonary vein isolation procedure, one of the pins in the navlink module was bent where the patch plugs in.A navlink was borrowed from another account which caused a 45 minute delay.The procedure was completed with no adverse consequences to the patient.
 
Manufacturer Narrative
One ensite velocity navlink module was received for evaluation.Visual inspection revealed the chest pin on the module was missing.The chassis and label were intact and had no signs of physical damage.No nicks or cuts were noted along the cable insulation.The reported event was able to be duplicated by visual inspection.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information and investigation provided to abbott, the reported event was confirmed, and the root cause was isolate to the missing pin on the module.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ NAVLINK MODULE
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16770977
MDR Text Key313597383
Report Number2184149-2023-00086
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH420995
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-