Model Number 174006
Device Problems Loss of or Failure to Bond (1068); Detachment of Device or Device Component (2907); Firing Problem (4011)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 03/22/2023
Event Type  Injury  
Event Description
According to the reporter, during a laparoscopic subtotal hysterectomy plus sacrocervicopexy procedure, when attaching the mesh to the sacrum, on three devices, the tacks were released poorly and with a lot of effort.The tacks also did not adhere to the fabric as desired.Additionally, the tackers came loose in the patient's cavity.It was noted that there was tissue damage at the stapling sites.To resolve the issue, a new tacker was used.
 
Manufacturer Narrative
D10 concomitant products: 174006, 174006 protack 5mm disp in (lot#p2k0027),174006, 174006 protack 5mm disp in (lot#p2k0027).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic subtotal hysterectomy plus sacrocervicopexy procedure, when attaching the mesh to the sacrum, on three devices, the tacks were released poorly and with a lot of effort.The tacks also did not adhere to the fabric as desired.Additionally, the tackers came loose in the patient's cavity and were retrieved with gasper.It was noted that there was tissue damage at the stapling sites.To resolve the issue, a new tacker was used.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTACK
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16779328
MDR Text Key313677535
Report Number2647580-2023-01621
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10884521070400
UDI-Public10884521070400
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K090470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number174006
Device Catalogue Number174006
Device Lot NumberP2K0027
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2023
Date Device Manufactured10/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age48 YR
Patient SexFemale
Patient Weight39 KG
-
-