MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-5
Device Problems No Display/Image (1183); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to an olympus repair facility, and an evaluation of the device was performed.During the evaluation, the customer's originally reported issue of an error code was not confirmed.The following additional findings were also noted: assorted chips, cracks, and scratches, wear of the angle wire resulting in angulation failing to meet the standard value, peeled video connector label, dirty components, damage on the lock lever preventing the bending section from being fixed firmly, and the number of broken wires in the bending tube blade exceeds the standard value.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
An olympus representative reported on behalf of the customer, that during preparation for use of the device in an unknown therapeutic procedure, their endoeye flex deflectable videoscope exhibited an unspecified communication error immediately after connecting.The procedure was completed using a similar replacement device.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event was caused by breakage or disconnection of the image sensor unit due to stress of repeated use, external factors or handling, or failure of the parts mounted on the electrical circuit board such as integrated circuit chips and capacitors.The events can be detected by following the instructions for use which state: "chapter 3 preparation and inspection, section 3.8 inspection of the endoscopic system.[inspection of the endoscopic image] confirm that the wli (white light imaging) and nbi (narrow band imaging) endoscopic images are normal.1.Before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2.Observe the palm of your hand in the wli and nbi endoscopic images.3.Confirm that light is output from the endoscope¿s distal end.4.Adjust the brightness level as appropriate.5.Confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities.6.Turn the angulation control levers slowly in each direction until it stops.7.Confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16788083
MDR Text Key313769644
Report Number9610595-2023-06498
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310355
UDI-Public04953170310355
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K201832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2023
Was the Report Sent to FDA? No
Date Manufacturer Received05/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-