ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751058
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2023
Event Type  malfunction  
Event Description
A physician reported that the cassette was leaking before the vitrectomy surgery.The surgery was completed by replacing the new product.There was no patient harm.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Current tracking indicates no adverse trend for this lot for this event.The returned sample was visually inspected and confirmed the green port of the cassette body cracked.A cassette pressure leak test was performed on the cassette and no leakage was detected within the cassette.A calibrated console representing the current software version was used to test the sample.Leakage occurred from the infusion port.The observed embrittlement on the port areas with fracture lines is consistent with some type of environmental stress cracking.We attempted to replicate the crack by hammering with a pin into the port, but it did not fail in a brittle manner.The root cause of the customer's complaint is potentially related to environmental stress cracking which would not have originated from the manufacturing process.After an investigation of this complaint, it has determined that no further actions will be pursued at this time as the root cause is not known.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16794577
MDR Text Key313810037
Report Number1644019-2023-00457
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657510580
UDI-Public00380657510580
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751058
Device Lot Number14MKV6
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-