Model Number D152
Device Problems Failure to Capture (1081); Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2023
Event Type  malfunction  
Event Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited suspected loss of capture (loc) on the right atrial (ra) lead.Further testing was recommended by boston scientific technical services (ts).Upon review, a signal artifact monitor (sam) episode was noted that resulted in the minute ventilation (mv) feature disabled.Additional information was received that a lead fracture was confirmed and the ra lead was explanted and replaced.This device was explanted due to therapy upgrade.No adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv) sensor that is related to a high impedance condition.Please see the description for more information regarding the specific circumstances of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN EL ICD DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16802449
MDR Text Key313923382
Report Number2124215-2023-19643
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526534324
UDI-Public00802526534324
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960040/S306
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/11/2019
Device Model NumberD152
Device Catalogue NumberD152
Device Lot Number538584
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
-
-