Model Number C0R47 |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/13/2023 |
Event Type
malfunction
|
Event Description
|
Name of procedure being performed: na (incoming inspection).Detailed description of event: [distributor] incoming inspection po #23094 this is a complaint from [distributor] evaluation team.The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: march 10, 2023.Please refer to the attachment file.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.We found the defects that already informed applied medical by cer.We just inform you the defect and return the products for following defects.You do not need investigation.Hairlike object in pouch: c0r47 lot #1477619 (1 ea.).Type of intervention: na (incoming inspection).Patient status: na (no patient involvement).
|
|
Manufacturer Narrative
|
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
|
|
Manufacturer Narrative
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of hair inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the hair likely originated from an operator during the manufacturing process.
|
|
Event Description
|
Name of procedure being performed: na (incoming inspection).Detailed description of event: [distributor] incoming inspection (b)(6).This is a complaint from [distributor] evaluation team.The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: march 10, 2023.Please refer to the attachment file.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.We found the defects that already informed applied medical by cer.We just inform you the defect and return the products for following defects.You do not need investigation.Hairlike object in pouch: c0r47 lot #1477619 (b)(4).Type of intervention: na (incoming inspection).Patient status: na (no patient involvement).
|
|