MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751613
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2023
Event Type  malfunction  
Event Description
A customer reported abnormal fluid/air exchange (fax) function, no air intake.There was no patient harm, and the surgery was completed after switching another one.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected and no obvious defects were found.It was noted the infusion cannula as was not returned.The non-invasive flow sensor (nifs) on the cassette housing was in good condition.A calibrated console representing the current software version was used to test the sample.The light-emitting diode (led) rings on the console turned green as the probe connectors were engaged to the console indicating the proper communication between the probe and the console.The ball in the check valve of the drip chamber moved freely per specification.The sample could prime and pass intraocular pressure (iop) calibration successfully.No anomalies were observed during priming.No system message code was generated during testing.Fluid flowed from the balance salt solution (bss) bottle to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test had completed.The infusion pressure was measured at multiple set points throughout the console range and met specifications.Toggling the infusion and the fluid/air exchange (f/ax) modes, fluid and air flowed from the cassette to the infusion line continuously without any bubble in various settings in all sub modes.With the infusion control on, fluid flowed from the cassette continuously without generating air to the infusion line.When the fluid/air exchange (f/ax) control was on, only air was introduced from the cassette to the infusion line.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.After investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.Based on our current tracking, there are no adverse trends for this reported complaint and lot.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16805850
MDR Text Key313907444
Report Number1644019-2023-00467
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657516131
UDI-Public00380657516131
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751613
Device Lot Number14Y6H4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received07/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/23/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-