WILSON-COOK MEDICAL INC INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE; PKL, LIGATOR, HEMORRHOIDAL Back to Search Results
Model Number G58010
Device Problems Separation Failure (2547); Failure to Fire (2610)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
During a colonoscopy, the physician used a cook instinct plus endoscopic clipping device.It was reported that the clip was attached to a postpolypectomy defect tissue and when the physician attempted to deploy the clip, it would not deploy.It remained attached to the drive wire.The clip had to be torn from the tissue and another of the same clipping device was used at the initial clipping site to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Pma/510(k): k212323.Investigation evaluation: the product said to be involved was returned in a biohazard bag.No lot number was returned with the device.Our laboratory evaluation of the product said to be involved confirmed the report based on the condition of the returned device.A visual examination confirmed the clip housing has separated from the coil catheter but remains attached to the end of the drive wire, in a closed position.A function test was attempted, in order to deploy the clip.With handle manipulation the clip deployed, with significant resistance encountered during deployment.The device was then advanced into the accessory channel of a pentax colonoscope (2.8mm channel) which was placed in a simulated lower gi position.The tip of the scope was retroflexed to simulate worst case scenario.With handle manipulation, the drive wire moved freely inside the outer sheath.A visual examination of the clip, catheter attachment, and coil catheter (distal end device components) showed no deformities or signs of damage.A product discrepancy or anomaly that could have contributed to the reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the returned device confirmed the report.The evaluation identified the clip housing separated from the coil catheter but remains attached to the end of the drive wire, this device condition is the cause of the deployment difficulty.The cause of the housing separation from the coil catheter is unknown.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.The instructions for use states, "note: if separation of clip is not immediate, gently move catheter back and forth or use other endoscopic maneuvers to separate catheter from clip.".Prior to distribution, all instinct plus endoscopic clipping devices are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE
Type of Device
PKL, LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key16827888
MDR Text Key314176188
Report Number1037905-2023-00190
Device Sequence Number1
Product Code PKL
UDI-Device Identifier00827002580107
UDI-Public(01)00827002580107(17)260103(10)W4670168
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG58010
Device Catalogue NumberINSC-P-7-230-S
Device Lot NumberW4670168
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OLYMPUS PCF-190
-
-