H10: the device was received for evaluation.During functional testing, the device was able to properly detect a downstream occlusion.A review of the event history log revealed 'downstream occlusion!' messages.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.The reported condition was verified.The cause of the condition was not determined.The force sensor will be replaced as a precautionary measure.Should additional relevant information become available, a supplemental report will be submitted.
|