Model Number 18320
Device Problem Insufficient Information (3190)
Patient Problem Hyperglycemia (1905)
Event Date 02/17/2023
Event Type  malfunction  
Event Description
It was reported, the patient's blood glucose rose to 19.4 mmol/l (349.2 mg/dl).The patient reported the pod alerted an occlusion alarm, while wearing the pod for between 25 and 36 hours on the abdomen.
 
Manufacturer Narrative
An inadequate seal between the o-ring and reservoir wall allowed fluid to pass by the o-ring and enter the device, causing corrosion on internal components.Corrosion on the leadscrew created drive resistance, resulting in the device generating an 0x14 alarm.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including confirming the blue soft cannula is inspected for any damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
thom mcnamara
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16832228
MDR Text Key314180513
Report Number3004464228-2023-10976
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385082000020
UDI-Public(01)20385082000020(11)220322(17)230922(10)PD1K03222221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211575
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/22/2023
Device Model Number18320
Device Catalogue NumberBLE-I1-529
Device Lot NumberPD1K03222221
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2023
Date Manufacturer Received04/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale
-
-