• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD DASH INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD DASH INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 18320
Device Problems Loss of or Failure to Bond (1068); Unintended Movement (3026)
Patient Problem Hyperglycemia (1905)
Event Date 04/19/2023
Event Type  malfunction  
Manufacturer Narrative
According to the complainant the device will not be returned for investigation.It was reported that the cannula had dislodged from the infusion site.This condition could interrupt insulin delivery and contribute to hyperglycemia.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Event Description
It was reported the patient's blood glucose level (bg) rose to 21 mmol/l (378 mg/dl) while wearing the pod between 5 and 24 hours.The pod reportedly fell off the infusion site (hip/buttocks), causing the cannula to dislodge.As treatment, a new pod was placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
thom mcnamara
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16832894
MDR Text Key314190489
Report Number3004464228-2023-11000
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385082000020
UDI-Public(01)20385082000020(11)220508(17)231108(10)PD1K05082221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211575
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/08/2023
Device Model Number18320
Device Catalogue NumberBLE-I1-529
Device Lot NumberPD1K05082221
Was Device Available for Evaluation? No
Date Manufacturer Received04/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/08/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age12 YR
Patient SexFemale
-
-