Model Number 479888
Device Problems High impedance (1291); Capturing Problem (2891); Impedance Problem (2950); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2023
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the left ventricular (lv) lead exhibited high, increasing thresholds and impedances.It was suggested that one of the lv lead electrodes may have calcification.The lv lead remains in use.No patient complications have been reported as a result of this event.
 
Event Description
It was further reported that the lv lead was reprogrammed.
 
Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance trend of the left ventricular pacing lead was rising.Analysis of the device memory indicated pacing capture threshold in the left ventricle was high.Analysis of the device memory indicated pacing capture threshold issue in the left ventricle.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN STABILITY QUAD MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16835780
MDR Text Key314263876
Report Number2649622-2023-10686
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/24/2020
Device Model Number479888
Device Catalogue Number479888
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6725 PIN PLUG, 5076 LEAD, W4TR04 CRTP,
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient SexMale
-
-