• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported that the ophthalmic console turned off on it owns.The details about the procedure and patient contact was not reported.
 
Manufacturer Narrative
The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The system was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16841139
MDR Text Key314264180
Report Number2028159-2023-00618
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657511501
UDI-Public00380657511501
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2011
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-