• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 19191
Device Problems Activation, Positioning or Separation Problem (2906); Material Twisted/Bent (2981)
Patient Problem Hyperglycemia (1905)
Event Date 04/17/2023
Event Type  malfunction  
Manufacturer Narrative
According to the complainant the device will not be returned for investigation.We are unable to confirm the reported needle mechanism failure, bent cannula or determine its root cause.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Event Description
It was reported that the pink slide insert did not move into the viewing window, which indicates a failure of the needle mechanism to deploy as intended during activation.The patient's blood glucose (bg) values reached over 250 mg/dl.The cannula was bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
thom mcnamara
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16841356
MDR Text Key314259479
Report Number3004464228-2023-11053
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385081120033
UDI-Public(01)20385081120033(11)210903(17)230303(10)L72310
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211575
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/03/2023
Device Model Number19191
Device Catalogue NumberZXP425
Device Lot NumberL72310
Was Device Available for Evaluation? No
Date Manufacturer Received04/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age26 YR
Patient SexFemale
-
-