|
A customer reported to olympus, a physician was using the thundebeat open fine claw 9 centimeter handpiece consumables and encountered some problems.The device started to alarm, and the surgeon realized the tip of the thunderbeat had dislodged.When the second thunderbeat also experienced this same event; the therapeutic haemorrhoidectomy was completed using a replacement device (harmonic).After the first event, the scrub nurse put the dislodged tip back to see if there was any missing part.It was confirmed that nothing went into the patient¿s body from either thunderbeats.There was no report of patient harm associated with this device.Related patient identifier: (b)(6).
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, , the falling of the tissue pad is likely occurred by the following mechanism: 1.Due to performing output while grasping nothing (including the case the user kept activating after cutting tissue), the distal end of the tissue pad was peeled away.2.The tissue pad was fell off because the pad added pulling load.The event can be detected/prevented by following the instructions for use which state: "do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation".Olympus will continue to monitor field performance for this device.
|