MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR PREMIERE50 ELECTRODE -EA; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR PREMIERE50 ELECTRODE -EA; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 227504
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a rotator cuff repair procedure, the vapr premiere50 electrode device was unable to coagulate or function.It was not reported if there were any delays to the surgical procedure.It was reported that a spare device was used to complete the surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: e1: the reporter¿s complete facility address was not provided.Investigation summary ¿ the product was not returned to j&j medtech orthopaedics, however a photo was provided for evaluation.Visual inspection of the returned photo found the device impregnated with foreign matter, presumably biological matter and sitting in a table, however no observations pertaining to the nature of the reported event could be observed.The overall complaint was not confirmed as the observed condition of the vapr premiere50 electrode -ea would have not contributed to the complained device issue.Based on the findings, the investigation could not draw any conclusions and no potential cause about the reported event can be established due to the insufficient evidence provided.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR PREMIERE50 ELECTRODE -EA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key22511605
Report Number1221934-2025-02586
Device Sequence Number12948013
Product Code GEI
UDI-Device Identifier10886705009923
UDI-Public(01)10886705009923(10)U2410003(17)271022
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K122425
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number227504
Device Lot NumberU2410003
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2025
Initial Report FDA Received Date07/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-