Event Date 05/10/2013
Event Type  Malfunction  
Event Description

On (b)(6) 2013 it was reported that the vns patient was interrogated and high impedance was found with an impedance value of greater than 10,000ohms. X-rays were requested and the device was turned off. It was stated that the lead appears intact but the pin may not be fully inserted and that the patient does not wish to have a further procedure to correct the problem at this time. Although surgery is likely, it has not occurred to date.

 
Manufacturer Narrative

Device failure is suspected, but did not cause or contribute to a death or serious injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLEAD MODEL UNKNOWN
Type of DeviceLEAD
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
nydia herzog
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key3144353
Report Number1644487-2013-01660
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 05/10/2013
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/04/2013
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received05/10/2013
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No