• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS INC PULSE GEN MODEL 102 GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS INC PULSE GEN MODEL 102 GENERATOR Back to Search Results
Model Number 102
Event Date 06/01/2013
Event Type  Death  
Event Description

On (b)(6) 2013 it was reported that the vns patient passed away on (b)(6) 2013 due to choking on a sandwich. Good faith attempts were made to physician but were not successful.

 
Manufacturer Narrative

 
Event Description

Additional information was received that the patient died of asphyxia by bolus of food. The generator and lead were returned to the manufacturer for evaluation. The generator output signal was monitored for more than 24-hrs, while the generator was placed in a simulated body temperature environment. Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period. The pulse generator diagnostics were as expected for the programmed parameters. In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications. There were no performance or any other type of adverse conditions found with the pulse generator. Note that the majority of the lead assembly (body) including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product. The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure. The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present. Continuity checks of the returned lead portion were performed, during the visual analysis, with no discontinuities identified. Based on the findings in the product analysis lab, there is no evidence to suggest an anomaly with the returned portion of the device. Note that since the majority of the lead assembly (body), including the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePULSE GEN MODEL 102
Type of DeviceGENERATOR
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key3195325
Report Number1644487-2013-01918
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/03/2013
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received06/28/2013
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator LAY USER/PATIENT
Device EXPIRATION Date11/30/2011
Device MODEL Number102
Device LOT Number014911
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer08/26/2013
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received08/19/2013
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 06/28/2013 Patient Sequence Number: 1
-
-