MAUDE Adverse Event Report: CYBERONICS PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problem Use of Incorrect Control/Treatment Settings (1126)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/17/2012
Event Type  malfunction  
Event Description
During review of the periodic programming history, it was discovered that a faulted diagnostic test occurred on (b)(6) 2013 and programmed the device to unintended settings.The device was interrogated prior to leaving the office; however, the settings were not corrected until (b)(6) 2013.
 
Manufacturer Narrative
Analysis of programming history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3551616
MDR Text Key4216992
Report Number1644487-2014-00059
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/05/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number857134
Other Device ID NumberVERSION 8.0
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received12/05/2013
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age65 YR
-
-