Model Number 302-20
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/04/2014
Event Type  malfunction  
Event Description
It was reported that device diagnostics resulted in high impedance (>10,000 ohms).There was no known patient manipulation or trauma that occurred that could have contributed to the high impedance.It was reported that the physician did not order x-rays and the device was not programmed off.The patient was referred to surgery.The patient underwent generator and lead replacement on (b)(6) 2014.It was reported that the explanting facility does not returned explanted devices to manufacturer for analysis; therefore analysis will not be performed.
 
Manufacturer Narrative
Device manufacturing records were reviewed.Review of manufacturing records of the lead confirmed all quality tests were passed prior to distribution.Device failure is suspected, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3663580
MDR Text Key4341687
Report Number1644487-2014-00616
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/30/2011
Device Model Number302-20
Device Lot Number200841
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received02/04/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/11/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age15 YR
-
-