The generator and lead were received for analysis.Analysis of the lead was completed on (b)(4) 2014.Note that a portion of the lead assembly including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis abraded openings were observed in several areas of the outer and inner silicone tubes.The positive connector ring quadfilar coil appeared to be broken approximately 21mm from the end of the connector boot.Scanning electron microscopy was performed and identified the area as having extensive pitting which prevented identification of the coil fracture type.A loose piece of broken negative connector pin quadfilar coil was observed inside the inner silicone tubing approximately 30mm from the end of the connector boot.Scanning electron microscopy was performed and identified the area as having evidence of being worn to the point of fracture with flat spots and pitting on the coil surface.The negative connector pin quadfilar coil appeared to be broken approximately 52mm from the end of the connector boot.Scanning electron microscopy was performed and identified the area as having evidence of being worn to the point of fracture with flat spots on the coil surface and no pitting.During the visual analysis of the returned 281mm portion quadfilar coil breaks were observed at 57mm from the area of the abraded opening, 85mm from the end of the abraded opening, and 117mm from the abraded opening.Scanning electron microscopy was performed and found that the area showed evidence of being worn to the point of fracture.Quadfilar coil 2 appeared to be broken approximately 145mm from the area of the abraded open outer / inner silicone tubes.Scanning electron microscopy was performed and identified the area as having evidence of being worn to the point of fracture with flat spots on the coil surface and no pitting.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.Low magnification sem analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution.What appeared to be white deposits were observed on the outer silicone tubing.Eds (energy dispersion spectroscopy ¿ provides chemical or element identity/composition analysis) was performed and identified the deposit as containing silicon, phosphorus, sodium and calcium.With the exception of the observed discontinuities the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no other discontinuities were identified.Based on the diagnostic results, a short-circuit condition appears to have existed.Analysis of the generator was completed on (b)(4) 2014.The generator performed according to functional specifications.During the product analysis there were no anomalies found with the pulse generator.
|