Analysis of the explanted generator and lead was completed.There were no adverse functional, mechanical, or visual issues identified with the returned generator.Analysis of the lead found abraded openings on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.Based on the findings, there is no evidence to suggest an anomaly with the returned portion of the device.Note that since a large portion of the lead assembly including the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.
|
It was reported that the vns patient underwent surgery on (b)(6) 2014 to explant her generator and lead.The explanted generator and lead have been returned to the manufacturer where analysis is currently underway.Further follow-up revealed that the patient¿s device had to be explanted due to generator and lead extrusion.The patient had been picking at her devices which caused them to extrude inferiorly and laterally to the original incision sites.The patient went to the er on (b)(6) 2014 due to device extrusion at the axillary incision site.Prior to explant, there was a noticeable breakdown of the patient¿s skin due to the extrusion.There were no signs of infection or abscess, but the patient was antibiotics as a precaution.The patient is expected to be re-implanted once her incision sites are healed.The patient has not been re-implanted to date.
|