REYNOSA MANUFACTURING LIBERTY CYCLER SET, SINGLE CONN./EXT. DL Back to Search Results
Catalog Number 050-87216
Device Problems Fluid/Blood Leak (1250); Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2014
Event Type  malfunction  
Event Description
A peritoneal dialysis patient reported dialysis solution leaked out of the cassette and into the cycler.Patient noticed the leak in the cassette compartment when removing the cassette.Patient's effluent remained clear and the patient had no adverse effects.
 
Manufacturer Narrative
The actual device was returned to the manufacturer for physical evaluation and the complaint is confirmed with a scratch on the cassette film.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE CONN./EXT. DL
Manufacturer (Section D)
REYNOSA MANUFACTURING
reynosa
MX 
Manufacturer (Section G)
REYNOSA PLANT
mike allen 1331
parque industrial reynosa
reynosa, tamaulipas, cp 8878 0
MX   88780
Manufacturer Contact
tanya taft, rn cnor
920 winter street
waltham, MA 02451
8006621237
MDR Report Key3950840
MDR Text Key4611353
Report Number8030665-2014-00473
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
PMA/PMN Number
K043363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Patient
Type of Report Initial
Report Date 05/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/01/2017
Device Catalogue Number050-87216
Device Lot Number14AR08031
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/09/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received05/16/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LIBERTY DIALYSIS CYCLER
-
-