CYBERONICS PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problems Use of Incorrect Control/Treatment Settings (1126); Failure of Device to Self-Test (2937)
Patient Problem Dysphasia (2195)
Event Date 06/27/2014
Event Type  malfunction  
Event Description
It was reported that the vns patient was having intolerable voice issues on an hourly basis.The patient¿s device was interrogated and found to be programmed to settings indicative of an interrupted system diagnostic test.The patient¿s device was tested during an office visit the previous day, but the system diagnostic test had to be interrupted because the test was too painful for the patient to tolerate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3952700
MDR Text Key4643606
Report Number1644487-2014-01853
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063829
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received06/28/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/26/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age45 YR
-
-