Model Number MODEL 250 |
Device Problems
Crack (1135); Failure to Power Up (1476)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/11/2014 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device failure occurred, but did not cause or contribute to death or serious injury.
|
|
Event Description
|
Analysis of returned tablet device was completed.A visual inspection of the tablet verified that the display was cracked.No anomalies associated with the software or battery was identified during the analysis.
|
|
Manufacturer Narrative
|
Evaluation codes: no device failure occurred that would contribute to the computer powering down during use.User interface caused the crack in the screen; however, the crack did not inhibit use of the computer.
|
|
Manufacturer Narrative
|
|
|
Event Description
|
It was reported that during surgery, the tablet kept shorting out and would power down while attempting device interrogation.A second programming tablet was used to successfully interrogate that patient's device.It was reported that the tablet had been accidentally dropped some time ago which caused a small crack in the screen, but this was not believed to cause the sudden power down.The programming tablet was received for analysis.Analysis is underway, but has not been completed to date.
|
|