MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Crack (1135); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/11/2014
Event Type  malfunction  
Manufacturer Narrative
Device failure occurred, but did not cause or contribute to death or serious injury.
 
Event Description
Analysis of returned tablet device was completed.A visual inspection of the tablet verified that the display was cracked.No anomalies associated with the software or battery was identified during the analysis.
 
Manufacturer Narrative
Evaluation codes: no device failure occurred that would contribute to the computer powering down during use.User interface caused the crack in the screen; however, the crack did not inhibit use of the computer.
 
Manufacturer Narrative
 
Event Description
It was reported that during surgery, the tablet kept shorting out and would power down while attempting device interrogation.A second programming tablet was used to successfully interrogate that patient's device.It was reported that the tablet had been accidentally dropped some time ago which caused a small crack in the screen, but this was not believed to cause the sudden power down.The programming tablet was received for analysis.Analysis is underway, but has not been completed to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3995031
MDR Text Key15956551
Report Number1644487-2014-01977
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 10.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received10/10/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age19 YR
-
-