• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS INC PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS INC PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Event Date 04/20/2011
Event Type  Malfunction  
Event Description

On (b)(6) 2014 during a review of programming history, it was observed that a faulted system diagnostic test occurred on (b)(6) 2011 which resulted in a settings change. While attempting to correct the settings, a partial programming event occurred which was corrected but the settings were still not fully corrected when they were changed as a result of the faulted system diagnostic test. The magnet output was not programmed back off as intended.

 
Manufacturer Narrative

Analysis of programming history.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROGRAMMING SOFTWARE
Type of DevicePROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4059021
Report Number1644487-2014-02199
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 08/08/2014
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received09/04/2014
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberMODEL 250
Device LOT Number523496
OTHER Device ID NumberVERSION 7.1
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received08/08/2014
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured01/31/2007
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-