• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Event Date 08/11/2014
Event Type  Malfunction  
Event Description

It was reported that the handheld device was not working properly. The handheld was returned to the manufacturer where analysis is currently underway.

 
Event Description

Analysis of the returned handheld device and software flashcard was completed. It was identified that the handheld was returned without a battery cover. As a result, the handheld would not power on. Once a known battery cover was installed, the handheld performed according to functional specifications. No anomalies associated with flashcard software or databases were identified during the flashcard analysis. The flashcard and software performed according to functional specifications.

 
Manufacturer Narrative

The handheld was returned without a battery cover. As a result, the handheld would not power on. Once a known battery cover was installed, the handheld performed according to functional specifications.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROGRAMMING SOFTWARE
Type of DevicePROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4077150
Report Number1644487-2014-02292
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial,Followup
Report Date 08/11/2014
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received09/10/2014
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberMODEL 250
Device LOT Number1073773
OTHER Device ID NumberVERSION 8.1
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer08/12/2014
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received09/29/2014
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured08/27/2009
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-