The user facility reported that a total of 22 pts had tested positive for e.Coli and pseudomonas aeruginosa, and candia after undergoing a bronchoscopy procedure from (b)(6) 2014 to (b)(6) 2014.All the pts were reported to have been examined using the same device but by different physicians.The user facility conducted two sets of culture test on the o-ring of the device using their in-house laboratory and a independent laboratory.Both culture tests were positive for micro-organisms.Many of the pts that tested positive for the aforementioned micro-organisms had pre-existing respiratory problems.The treatment provided to the pts varied, but many received iv antibiotics.All pts have fully recovered and have been cleared of the suspected infections.The user facility has removed the device from use as of (b)(6) 2014.
|
The device referenced in this report has not yet been returned to olympus for eval.The exact cause of the pts outcome could not be conclusively determined at this time.However, review of the instrument history showed that the device was purchased on (b)(4)2013, but the device has not been returned to olympus for service since then.If additional info becomes available at a later time, this report will be supplemented.Please reference the associated devices, see mfr report # 2951238-2014-00400, 00401, 00402, 00403, 00404, 00405, 00406, 00407, 00408, 00409, 00410, 00411, 00412, 00413, 00414, 00415, 00416, 00417, 00418, 00419, 00420.
|
This supplemental report is being submitted to provide additional information regarding the in-service visits provided by the endoscopy support specialist (ess) on august 21, 2014 and october 3, 2014.The ess visited the user facility to observe the user facility's reprocessing practices.During the in-service visit it was observed that the endoscopes were being hung with peel pack covers (distal tip protectors) over the insertion tubes while drying.As part of educating the customer, a copy of the distal tip protector letter from infection control was provided to the user facility as well as both manuals on the proper cleaning and reprocessing of the 180 series bronchoscopes and colonoscopes.It was also reported that no other reprocessing deviations were found during the on-site visit.Olympus does not recommend nor endorse the use of the distal tip protectors during storage.It was noted that the endoscope distal tip protectors are not to be reprocessed or reused.Reuse could present an infection control risk if the distal end protectors cannot be thoroughly dried.Additionally, the attachment of a non-sterile distal tip protector to the endoscope which has been high-level disinfected or sterilized may cause recontamination of the endoscope.Olympus bronchoscope instruction manual advises users: "storage of endoscope: to hang the endoscope in the storage cabinet with the distal end hanging freely." if additional or significant information is available at a later time, this report will be supplemented accordingly.
|