CYBERONICS, INC. LEAD MODEL 302 Back to Search Results
Model Number 302-30
Event Date 08/14/2014
Event Type  Malfunction  
Event Description

Clinic notes dated (b)(6) 2014 note that interrogation of the device showed high impedance and was verified with a second interrogation. The patient was referred for surgery. No known surgical intervention has been performed to date. No additional relevant information has been received to date.

 
Event Description

Analysis of the lead was completed on 12/11/2014. Note that the electrodes were not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product. During the visual analysis the (-) connector pin quadfilar coil appeared to be broken approximately 2. 5mm from the electrode bifurcation (and the coil appeared to be kinked). The (+) connector ring quadfilar coil appeared to be broken approximately 3mm from the electrode bifurcation. Scanning electron microscopy was performed on the (-) connector pin quadfilar coil break (found at 2. 5mm) and identified as having evidence of a stress induced fracture (fatigue appearance) with mechanical damage and pitting. Pitting was observed on the coil surface. Scanning electron microscopy was performed on the connector end of the (+) connector ring quadfilar coil break (found at 3mm) and identified the area as having evidence of a stress induced fracture (fatigue appearance) with mechanical damage and fine pitting. It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting. Low magnification sem analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution. The abraded opening found on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing. What appeared to be white deposits were observed on the outer silicone tubing. Eds (energy dispersion spectroscopy ¿ provides chemical or element identity/composition analysis) was performed and identified the deposit as containing silicon, phosphorus, sodium, magnesium, aluminum, sulphur and calcium. With the exception of the observed discontinuities the condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure. No other obvious anomalies were noted. The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present. Continuity checks of the returned lead portion were performed, during the visual analysis, and no other discontinuities were identified.

 
Event Description

It was reported that the patient underwent generator and lead replacement. The generator and lead were received for analysis. Generator analysis was completed on 12/08/2014. A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications. There were no performance or any other type of adverse conditions found with the pulse generator. Analysis of the lead is underway, but has not been completed to date.

 
Manufacturer Narrative

Device failure is suspected, but did not cause or contribute to a death or serious injury.

 
Manufacturer Narrative

Device failure occurred, but did not cause or contribute to a death or serious injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLEAD MODEL 302
Type of DeviceLEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4106898
Report Number1644487-2014-02419
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 08/25/2014
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received09/22/2014
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date07/31/2010
Device MODEL Number302-30
Device LOT Number200556
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer11/18/2014
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received12/11/2014
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured07/12/2007
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-