Model Number MODEL 250
Event Date 06/01/2014
Event Type  Malfunction  
Event Description

Analysis of a returned handheld was completed on 09/18/2014. During the analysis, it was identified that the handheld would not power on. The cause for the anomaly is associated with a swollen main battery. The swollen battery bent the battery cover causing the battery latch switch to register that the latch was unlocked, thus preventing the handheld from powering on. No further anomalies were identified. Analysis of the flashcard was completed on 09/18/2014. No anomalies associated with flashcard software or databases were identified during the flashcard analysis. The flashcard and software performed according to functional specifications.

 
Manufacturer Narrative

Device failure occurred, but did not cause or contribute to a death or serious injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROGRAMMING SOFTWARE
Type of DevicePROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4171215
Report Number1644487-2014-02686
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 09/18/2014
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/14/2014
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberMODEL 250
Device LOT Number1077281
OTHER Device ID NumberVERSION 8.1
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Event Location Other
Date Manufacturer Received09/18/2014
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured10/08/2009
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-