Model Number MODEL 250
Event Date 09/23/2014
Event Type  Malfunction  
Event Description

Analysis of the returned handheld device and software flashcard was completed. It was identified that the handheld was unable to establish communication with a known good wand and generator. The cause for the anomaly is associated with a broken wire connection in the serial cable. Once the wire was soldered onto the serial cable pcb, no further anomalies were identified. The handheld did not shut-down or power-off by itself during the analysis. No anomalies associated with flashcard software or databases were identified during the flashcard analysis. The flashcard and software performed according to functional specifications.

 
Manufacturer Narrative

Device failure occurred, but did not cause or contribute to death or serious injury.

 
Event Description

It was reported that the physician's handheld was not working properly. It was reported that the handheld was powering down without being prompted. Troubleshooting was performed, but did not resolve the issue. A new programming tablet was provided to the physician. The handheld was received for analysis. Analysis is underway, but has not been completed to date.

 
Manufacturer Narrative

Device manufacturing records were reviewed. Review of manufacturing records of the handheld confirmed all quality tests were passed prior to distribution.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROGRAMMING SOFTWARE
Type of DevicePROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4178457
Report Number1644487-2014-02722
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation PHYSICIAN ASSISTANT
Type of Report Initial,Followup
Report Date 09/23/2014
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/16/2014
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberMODEL 250
Device LOT Number1081704
OTHER Device ID NumberVERSION 8.1
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer09/26/2014
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received10/20/2014
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured08/31/2005
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-