Model Number MODEL 250
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/22/2010
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
During review of programming and diagnostic history, it was observed that an interrupted system diagnostic test occurred that caused an unintended change in device settings during an office visit on (b)(6) 2010.The physician corrected the settings; however, the magnet pulse width was not corrected.No patient adverse events were reported.
 
Manufacturer Narrative
Review of the available programming and diagnostic history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4200277
MDR Text Key12583633
Report Number1644487-2014-02822
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/11/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 7.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received11/26/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age7 YR
-
-