• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS INC BIPOL LEAD MODEL 300

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS INC BIPOL LEAD MODEL 300 Back to Search Results
Model Number 300-20
Device Problems Break (1069); Device Inoperable (1663); Moisture or Humidity Problem (2986); Output Problem (3005)
Patient Problem No Code Available (3191)
Event Date 10/23/2014
Event Type  malfunction  
Event Description
It was reported that the vns patient was scheduled for surgery due to end of service.Pre-operative diagnostic results showed output current = low.During the procedure, the outer silicone tubing of the lead appeared to be ¿stripped¿ and fluid was noted to have entered the outer silicone tubing.The surgeon elected to wrap catheter tubing around the stripped portion of the lead.Only the generator was replaced during the procedure.Post-operative diagnostic results showed output current = ok.The explanted generator has not been returned to date.
 
Manufacturer Narrative
Device failure is suspected, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOL LEAD MODEL 300
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4262831
MDR Text Key5010875
Report Number1644487-2014-03085
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2003
Device Model Number300-20
Device Lot Number4398
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received10/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2001
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
-
-