Analysis of the lead was completed on 12/15/2014.Note that the lead assembly (body) including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.What appeared to be white deposits were observed in various locations.Eds (energy dispersion spectroscopy ¿ provides chemical or element identity/composition analysis) was performed and identified the deposit as containing silicon, phosphorus, sodium, sulphur and calcium.The condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The setscrew marks found on the lead connector pins provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no discontinuities were identified.Analysis of the generator was completed on 12/16/2014.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no performance or any other type of adverse conditions found with the pulse generator.
|