CYBERONICS, INC. BIPOL LEAD MODEL 300 Back to Search Results
Model Number 300-20
Event Date 12/08/2014
Event Type  Malfunction  
Manufacturer Narrative

Evaluation code conclusion code, corrected data: the initial report inadvertently reported the incorrect code.

 
Event Description

It was reported that during generator replacement for near end of service (neos=yes), the surgeon observed a possible abraded opening in the outer tubing of the patient¿s existing lead near the generator pocket after the generator was replaced. When the observance was reported to the manufacturer field representative in the operating room, the new generator was in the half-closed pocket and diagnostics had been performed. The surgeon elected to not to replace the lead, as the patient was not experiencing any side effects prior to surgery, has been receiving great results from the vns therapy, and above all because the impedance was within normal limits.

 
Manufacturer Narrative

Manufacturer device history records were reviewed. Review of the lead device history records confirmed all quality specifications were passed prior to distribution with no non-conformances noted. Device failure is suspected, but did not cause or contribute to a death or serious injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBIPOL LEAD MODEL 300
Type of DeviceLEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4336631
Report Number1644487-2014-03367
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial,Followup
Report Date 12/08/2014
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/17/2014
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date04/30/2003
Device MODEL Number300-20
Device LOT Number2837
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received12/17/2014
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured05/01/2001
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-