Model Number MODEL 100
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2014
Event Type  malfunction  
Event Description
It was reported that the autopulse platform's display did not work.No adverse patient sequelae was reported.No further information was provided.
 
Manufacturer Narrative
Zoll has not yet received the autopulse platform in complaint for investigation.A supplemental report will be filed if and when the product is returned and investigation has been completed.
 
Manufacturer Narrative
The autopulse platform in complaint was returned to zoll on 01/08/2015 for investigation.Investigation results as follows: visual inspection of the returned platform was performed and found that the battery lock clip was broken and the lcd display background was not functioning properly.During initial functional testing, the system was turned on/off with no problems and ran using a mannequin with no anomalies or errors exhibited.A review of the autopulse platform's archive was performed and found no errors to have occurred on the reported event date of (b)(6) 2014.Based on the investigation, the processor pca was replaced to remedy the customer's reported complaint of the platform's display background not working.The other damaged part noted during visual inspection of the returned platform was also replaced.The platform underwent and passed all final functional testing.In summary, the customer's reported complaint of the platform's display background not working was confirmed through visual inspection of the returned platform.The root cause was determined to be a defective processor pca, which was replaced, remedying the issue and allowing the platform to function as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key4385445
MDR Text Key5361787
Report Number3010617000-2015-00012
Device Sequence Number1
Product Code DRM
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0740-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received02/25/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-