It was reported that the physician's programming system was unable to interrogate a known working generator.Troubleshooting was performed which identified that the physician's programming system was found to not interrogate.The physician was provided a new programming system and the handheld and wand were returned for analysis.Analysis of the handheld was completed on (b)(4) 2014.During the analysis an intermittent power connection associated with the serial cable was identified.The cause for the anomaly is associated with a bent serial cable plug leaf spring that prevented electrical contact with the ac adapter barrel connector.The cause for the bent spring is unknown, but is most likely associated with mishandling of the device.Analysis of the flashcard was completed on (b)(4) 2014.No anomalies associated with flashcard software were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.Analysis of the wand was completed on (b)(4) 2014.The returned battery was depleted and was not the recommended type per labeling (battery type labeling recommends alkaline).After the battery was substituted, the programming wand performed according to functional specifications.
|