MAUDE Adverse Event Report: BAXTER HEALTHCARE - SINGAPORE HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SINGAPORE HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 01/06/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The homechoice device was returned and evaluated by the product analysis lab (pal).A device history record review revealed no issues that could have caused or contributed to the reported issue.An internal/external inspection was performed and the device passed, along with all of the electrical testing.During the evaluation, the fluid was transferred outside of the returned instrument testing evaluation (rite) specified limits and the volumetric accuracy test that followed, showed that the device had failed.Pal performed a more detailed inspection of the door assembly.The inspection showed that the piston foam was deteriorated.The results of the evaluation revealed the cause of the failure to be the deteriorated piston foam.The piston foam was to be scrapped and the device was sent for servicing.A capa has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned homechoice device, a baxter technician determined the device failed fluid volume accuracy testing.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park
singapore 2573
SN  2573
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park
singapore 2573
SN   2573
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4460888
MDR Text Key20250906
Report Number1416980-2015-03529
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C8310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received02/04/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-