• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. EVERFLO OXYGEN CONCENTRATOR; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. EVERFLO OXYGEN CONCENTRATOR; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 102000
Device Problem Leak/Splash (1354)
Patient Problem Low Oxygen Saturation (2477)
Event Date 01/07/2015
Event Type  Injury  
Event Description
The manufacturer received info alleging a patient's blood oxygen level was low while on an everflo oxygen concentrator.The patient was instructed by the durable medical equipment (dme) supplier to use their back up oxygen supply.There was no medical intervention required by the patient.
 
Manufacturer Narrative
The device was evaluated by a third party service center.The technician found the sieve canister to be leaking and the filters dirty.The sieve canister and filters were replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO OXYGEN CONCENTRATOR
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
kennesaw GA
Manufacturer Contact
don mcandrews
1740 golden mile highway
monroeville, PA 15146
7243873965
MDR Report Key4501533
MDR Text Key19765772
Report Number1040777-2015-00004
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061261
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number102000
Device Catalogue Number102000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/07/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age86 YR
-
-