Model Number 302-20
Event Date 06/05/2013
Event Type  Malfunction  
Event Description

During review of programming and diagnostic history, it was observed that the vns patient¿s device was tested during an office visit on (b)(6) 2013 and system diagnostic results revealed high impedance (dc dc ¿ 7) and an end of service condition. No patient adverse events were reported. An implant card was received indicating that the patient underwent generator and lead replacement surgery on (b)(6) 2013. The explanted devices have not been returned to date.

 
Manufacturer Narrative

Review of the available programming and diagnostic history. Device manufacturing records were reviewed. Review of manufacturing records confirmed that the lead passed all functional tests prior to distribution. Device failure is suspected, but did not cause or contribute to a death or serious injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLEAD MODEL 302
Type of DeviceLEAD
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4524697
Report Number1644487-2015-03883
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 01/23/2015
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received02/18/2015
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date11/30/2009
Device MODEL Number302-20
Device LOT Number200371
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Event Location Other
Date Manufacturer Received01/23/2015
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured11/20/2006
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-