It was reported that during generator replacement surgery device diagnostics with the new generator and existing lead were within normal limits both inside and outside of the generator pocket and once the incision was closed device diagnostics showed ifi - yes.It was reported that electrocautery was not brought back into the field during incision closure; however, it was on the table near the patient which may have discharged during incision closure.A new generator was then implanted.A device programming decoder showed that the battery voltage changed from 3.354v to 2.022v between 9:04:31 and 10:00:41 on (b)(6) 2015.The generator was received for analysis.Analysis is underway, but has not been completed to date.
|
Analysis of the returned generator was completed.An end-of-service warning message was verified and found to be associated with the output being disabled by the pulse generator.Burn marks were observed on the pulse generator case, which indicated that the pulse generator may have been exposed to an electro-cautery tool during device implant or explant.A reset of the pulsedisable bit in the generator memory was performed to allow for an output to once again be provided by the generator for subsequent testing.The battery of the returned generator was measured as 3.095 volts showing an ifi=no condition.The downloaded data revealed that 1.629% of the battery had been consumed.Other than the observed pulse disablement and burn marks, there were no performance or any other type of adverse conditions found with the pulse generator.
|