Model Number 302-20
Event Date 01/30/2015
Event Type  Malfunction  
Event Description

Further information was received indicating that manipulation of the devices or trauma was not suspected. The generator was disabled. No increase of the seizure rate was noted.

 
Event Description

It was reported that device diagnostics resulted in high impedance. X-rays were taken and sent to manufacturer for review. Based on the x-rays images, no obvious cause of high lead could be found. Attempts to obtain additional relevant information have been unsuccessful to date. No known surgical intervention has been performed to date.

 
Manufacturer Narrative

Device failure is suspected, but did not cause or contribute to a death or serious injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLEAD MODEL 302
Type of DeviceLEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4555396
Report Number1644487-2015-04014
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/30/2015
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received02/27/2015
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date12/31/2016
Device MODEL Number302-20
Device LOT Number3490
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received03/04/2015
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured12/11/2012
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-