Event Date 02/03/2015
Event Type  Malfunction  
Event Description

It was reported that device diagnostics resulted in high impedance (>10,000 ohms). The patient was referred for surgery. There was no known patient manipulation or trauma that may have caused or contributed to the high impedance. X-rays were sent to manufacturer for review. Review of x-rays did not identify any obvious discontinuities with the lead. It was observed that the lead pin may not be fully inserted into the generator header. No known surgical interventions have been performed to date.

 
Manufacturer Narrative

Device failure is suspected, but did not cause or contribute to a death or serious injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLEAD MODEL UNKNOWN
Type of DeviceLEAD
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4570834
Report Number1644487-2015-04044
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 02/03/2015
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received03/04/2015
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received02/03/2015
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-