Device Problem
Display or Visual Feedback Problem (1184)
Patient Problem
No Known Impact Or Consequence To Patient (2692)
Event Date 08/13/2010
Event Type
Malfunction
Event Description
It was reported the module is producing ecgs (electrocardiograms) that are too small or too tall and narrow to be readable.
The ecg module was returned for exchange/swap.
No patient complications have been reported as a result of this event.
Manufacturer Narrative
This event was assessed and is being reported as a part of a retrospective review of events, which was in response to mdr criteria revisions and on-going process improvements.
Product event summary: analysis found the device had contamination in the usb (universal serial bus) connectors which was determined to be a potential cause for the reported event.
(b)(4).