Model Number RT380 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
A hospital in the (b)(6) reported that swivel wye of an rt380 breathing circuit is disconnecting easily.No patient consequence reported.
|
|
Manufacturer Narrative
|
(b)(4).The complaint rt380 evaqua2 breathing circuit is en route to fisher & paykel healthcare (b)(4) for evaluation.A follow up report will be provided upon completion of our investigation.
|
|
Manufacturer Narrative
|
(b)(4).Method: the complaint rt380 evaqua2 breathing circuit was returned to fisher & paykel healthcare (b)(4) for evaluation.The device was visually inspected and pressure tested.Result: the visual inspection did not reveal any damage to the breathing circuit of the dryline.The pressure test was performed and the product was within specification.No fault was found with the device.Conclusion: we were unable to determine the cause of the reported incident as no fault was found with the device.All rt380 breathing circuits are visually inspected and pressure tested for leaks before releasing for distribution.Any breathing circuit which fails any of these tests is discarded.The user instructions supplied with the rt380 breathing circuit state: perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.The hospital staff correctly checked the breathing circuit before patient use, which is in line with our user instructions.
|
|
Event Description
|
A hospital in the (b)(6) reported that swivel wye of an rt380 breathing circuit is disconnecting easily.No patient consequence reported.
|
|
Search Alerts/Recalls
|