Catalog Number 700-3345
Device Problems False Negative Result (1225); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 03/12/2015
Event Type  malfunction  
Event Description
Short sample detector is not detecting an empty tube run on the instrument.
 
Manufacturer Narrative
During a customer visit, an iris field service engineer (fse) ran a pt rack containing empty tubes and observed that the short sample detector did not flag some of the empty tubes and produced a false negative result.The fse adjusted the short sample detector to correct the issue and when the same rack was rerun, the instrument read all tubes and flagged accordingly.There were no patients involved in this troubleshooting exercise.The customer instrument was operational.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ200
Type of Device
AUTOMATED URINE MICROSCOPY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton ave.
chatsworth, CA 91311
8185277379
MDR Report Key4688786
MDR Text Key20772339
Report Number2023446-2015-00123
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K022774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-3345
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/12/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-