The complaint device was returned to stryker sustainability solutions (sss) for evaluation.Inspection of the returned device revealed biological material on the distal tip and an indention in the teflon pad.The device was actuated multiple times and confirmed to have unacceptable mechanical functionality.The jaw would stick in the close position and not open completely.The device was examined further and a build up of biological material was found on the distal tip.The device tip was cleaned, and jaw actuation was re-examined; the jaw was able to open/close without sticking.The device was then connected to a scalpel generator and appropriate hand piece.The scalpel was tested and gave an error code 5.The rod was inspected for fractures, and a crack in the blade was observed.A review of the device history record indicates the device met all inspection and test criteria prior to release.Therefore, the most likely root cause of the cracked blade is the blade contacting a hard object, possibly a staple or surgical clip, during clinical use.The instructions for use for harmonic ace curved shears state: "avoid contact with any and all metal or plastic instruments or objects during instrument activation.Contact with staples, clips, or other instruments during instrument activation may result in premature blade failure, resulting in generator solid tone or instrument error.¿.
|