|
Model Number HAR36 |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/11/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that during the procedure the generator of the ultrasonic scalpel alarmed, "relax pressure on blade." the alarm occurred three times, and the facility examined and reset the device each time.There was no medical intervention, surgical delay, or adverse consequences to the patient.The procedure was completed successfully.
|
|
Manufacturer Narrative
|
The complaint device was returned to stryker sustainability solutions for evaluation.Inspection of the returned device revealed evidence of clinical use including biological material on the distal tip and an indention in the teflon pad.The shaft rotation and the jaw actuation of the device were found to be acceptable.The device was connected to an generator using the appropriate hand piece.The scalpel was tested and failed with two "tighten assembly" errors (after which the assembly was tightened) and a "replace instrument" error.The device was examined further, and a build up of dried fluid was noticed along the rod.The distal gasket on the rod was examined and revealed signs of damage and wearing of the gasket lip.The rod was inspected for fractures, and a crack as well as gouges in the blade were observed.A review of the device history record indicates the device met all inspection and test criteria prior to release.The most likely root cause of a cracked blade is activating the device on a surgical clip or staple.The reported event will continue to be monitored through post-market surveillance.The instructions for use state: "avoid contact with any and all metal or plastic instruments or objects when the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.".
|
|
Search Alerts/Recalls
|
|
|